AELAB Isocratic & Binary Gradient System HPLC LC1620AFeatures:+ Pumping system: Isocratic system and Binary Gradient system.
+ Different detectors available: UV, ELSD, FLD, RID.
+ UV Detector for routine analysis/ RID for universal detection in isocratic analysis, especially necessary in GPC system.
+ Superior high-pressure pump achieves accurate solvent delivery with high-precision and low-fluctuation.
+ Advanced and innovative design modules provide high reliability and accuracy for long time operation.
+ The pump works in-parallel and ensures accuracy and duration.
+ Electronic signature function.Technical Specification : Model | LC1620A |
Isocratic system | Binary Gradient system |
LC1620A Pump | Specifications: |
Delivery system | Dual-plunger in series, Float plunger design, Single stroke discharged fluid volume is 30 ul |
Flow rate range | 0.001-9.999mL/min Increment: 0.001ml/min |
Flow rate accuracy | ≤±0.5% or 10 µL/min (1mL/min, water, room temperature) |
Flow rate precision | RSD≤0.1% (1mL/min, water, room temperature) |
Pressure fluctuation | ≤ 0.5% |
Pressure range | ≤ 42 Mpa for 316L pump head ≤ 30 Mpa for PEEK pump head |
Pipeline connection | 1/16” standard pipeline connection |
Display parameters | OLED graphic display |
Control | Instrument panel control or computer anti-control |
Power supply | 100~240VAC, power: 75W |
LC1620A UV Detector | Specifications: |
Lamp | Deuterium or halogen lamp selectable |
Detection system | Single or dual-wavelength mode selectable |
Wavelength range | 190-700nm Increment: 1nm |
Wavelength accuracy | 1nm |
Wavelength precision | ±0.1nm |
Volume of flow cell | 10μl |
Detect range | 0-2Au |
Baseline noise | ±0.5×10-5 AU |
Baseline drift | 2×10-4 AU |
Detection Limit | 4×10-9 g/ml(Naphthalene/methyl alcohol solvent) |
Pipeline connection | 1/16” standard pipeline connection |
Display parameters | OLED graphic display |
Control | Instrument panel control or computer anti-control |
Power supply | 100~240VAC, power: 80W |